K-Patents SeedMaster 2 User Manual

Page 84

Advertising
background image

11 COMMUNICATION

84

84

2. XOR CRC high with the 1

st

data byte, the result remains in the CRC word.

3. Shift CRC 1 bit right.

4. If the most significant bit of the CRC was 0 before the shift, the next step is step 3 (go to step 3).

5. If the least significant bit of the CRC was 1 before the shift (1 was shifted out) then XOR CRC and the

CRC calculation polynomial A001 (hex), the result remains in the CRC word.

6. Repeat steps 3 & 4 until 8 shifts.

7. XOR CRC and the next data byte, the result remains in the CRC word.

8. Repeat 3-7 steps with all data bytes.

9. The result is the CRC code to be sent or verified.

The implemented MODBUS commands:

Read multiple MODBUS registers command, function code : 03(03h):

Query:

A

F

RAH

RAL

RNH

RNL

CH

CL

Response:

A

F

BN

D1

DN

CH

CL

Write multiple MODBUS registers command, function code : 16 (10h):

Query:

A

F

RA

H

RA

L

RN

H

RN

L

BN

D1

CH

CL

Response:

A

F

RAH

RAL

RNH

RNL

CH

CL

where

A

– slave address

F

– function code

RAH, RAL – starting address of the MODBUS registers, order: high byte first

RNH, RNL – number of the MODBUS registers, order: high byte first

BN

– number of the data bytes sent (0..255)

D1….Dn

– n data bytes, n=BN*2, order: high byte first

CH, CL

– CRC control code, order: high byte first

Examples

Function code: 3 (read), RTU mode

Read the first 32 data (64 registers, i. e. 128 bytes) from the „read data” table of the 1

st

Crystallizer of

SeedMaster 2 having the station address 17 (11h):

Query:

Slave

address

Function

code

1

st

register

address

(high byte)

1

st

register

address

(low byte)

No of

registers

(high byte)

No of

registers

(low byte)

CRC
code

(2 bytes)

11h

03h

00h

00h

00h

40h

CRC


Response:

Slave

address

Function

code

No of the

bytes sent

Data bytes

CRC

(2 bytes)