8 SET UP SeedMaster 2

65

65

SET UP -> ANALOG OUTPUT ->

1 / 0 . 1

SET UP -> AN. OUTPUT -> SUPERSAT.

1 / 0 . 1

CH . CURR. mA

RANGE

CLEAR

AO_2

4 - 20

0 . 7 - 2

STANDARD CURRENT

4 - 20 mA

DENSITY :

RANGE

0 . 7 - 2

MA . SOLIDS C. :

LOW LIMIT

0 . 9

CRYST. CONT

. : AO_1

4 - 20

0 -60

HIGH LIMIT

1 . 1 8

MEAN CR. SIZE :

CHANNEL

AO_2

CONSISTENCY :

ENGINEERING UNIT

--

MO. LI. PURITY :

MO. LI. CONC. :

MOTOR CONS. :

CHANGE

BACK

ACCEPT

CHANGE

BACK

SUPERSATUR. :

a) b)

Fig. 8.12

Figure 8.12 a) shows how to set up an output (SUPERSATURATION) channel.

To modify
type CHANGE and continue the Set up as shown in Fig. 8.12 b) and Fig. 8.13.

To free
an already reserved output channel select item (Fig. 8.12 a)), then select CLEAR and type CHANGE
(Fig. 8.12 b)).

To add
a new output select item (for example: DENSITY) and type CHANGE (Fig. 8.12 a)).
Select CHANNEL and type CHANGE (Fig. 8.12 b)). Available free channels will be offered automatically. If
there is no available channel, make one free first.

SET UP -> AN . OUT -> SUPERS. -> RANGE

1 / 0 . 1

MIN.

MAX.

CURRENT VALUE :

0,7

2

NEW VALUE :

>

0,7

1,8

CHANGE

BACK

Fig. 8.13

IMPORTANT NOTE:
Having completed selections or changes during any SET UP operation it is highly recommended to
check if the final settings are really correct (go to SET UP -> ANALOG OUTPUT). Do not forget to
close the operations by pushing the ACCEPT or ENTER key.





9

MANUAL SEEDING

It is highly recommended to practice automatic seeding whenever possible. It will result in reproducible
strikes, which, if the set-point for seeding was selected properly (see: Ch. 2.5) and the process of
crystallization was also controlled well, will provide consequently high-quality product.
Manual seeding can be the selected (though not recommended) mode of seeding. However, in un-common
situations manual seeding of the crystallizer might be necessary. The mode of seeding (MANUAL SEED, or
AUTO SEED) can and should be configured in advance during CONFIGURE -> SEEDING (see: Ch. 7.5).

Advertising