View menu – Wavecom W61PC V7.5.0 User Manual

Page 64

Advertising
background image

54

First start

WAVECOM Decoder W61PC/LAN Manual V7.5

Note: The full range of the“Fine Tune” process is 0.1 % deviation. For deviations larger than this use “Fine
Tune” repeatedly until the desired effect has been achieved.

If you succeeded in correcting the sampling rate, i.e. the vertical lines have been straightened and are
perpendicular to the time axis, press “OK” to save the measured value.

As the sampling rate calibration feature utilizes an AM demodulator, input level should be as high as pos-
sible, resulting in a high level of contrast which facilitates the adjustments. If necessary use the “Demodu-
lator” menu to adjust the AM gain and offset.

View Menu