The trigger menu – Tektronix AWG610 User Manual

Page 157

Advertising
background image

The Setup Menu Screen

AWG610 Arbitrary Waveform Generator User Manual

3-39

The Trigger Menu

The Trigger menu lets you set instrument external signal trigger parameters. The
Trigger menu commands are Source, Slope, Level, Impedance, and Interval.

This button lets you set the instrument trigger source. You can select either
External or Internal.

To set the trigger signal source, push SETUP (front)

!Trigger (bottom)!

Source (side) to toggle between External and Internal.

If you select External, the instrument uses the signal connected to the rear-panel
TRIG IN connector. The external trigger signal must meet the following
requirements:

Table 3-5: External trigger signal requirements

Requirement

Values

Input voltage range

±

10 V into 1 kW impedance

±

5 V into 50 W

impedance

Minimum pulse width

10 ns

If you select Internal, the trigger signal generated in the instrument will be used.
For the internal trigger, you can set only the trigger interval. In the Gated mode,
the internal trigger does not work.

This button lets you set the external trigger signal slope (or polarity in the Gated
run mode) on which to trigger the instrument. You can specify to trigger on the
rising (Positive) or falling (Negative) edge.

To set the trigger slope, push SETUP (front)

!Trigger (bottom)!Slope (side)

to toggle between Positive and Negative.

NOTE. In the Gated run mode, triggering occurs for the time period that the
external trigger signal level is greater than or equal to the specified trigger level
setting.

Source

Slope (or Polarity)