Event input and enhanced mode tests – Tektronix AWG610 User Manual

Page 405

Advertising
background image

Appendix B: Performance Verification

AWG610 Arbitrary Waveform Generator User Manual

B-45

Event Input and Enhanced Mode Tests

These procedures check the event input signals and enhanced mode operation.

NOTE. The event input check with strobe off and the strobe input check are
structured as a continuous test. After
Check Event Input with Strobe Off, the
next test uses the connections and oscilloscope settings from the previous test.

Equipment

required

AĂ50ĂW SMAĂcoaxial cable,ĂaĂSMA(Fe)ĆBNC(Ma)Ăadapter an

oscilloscopeĂ(TDS700), and customĆmade ground closure. See Figure

B-3 for the connections.

Prerequisites

The AWG610 Arbitrary Waveform Generator must meet the

prerequisites listed on page B-8.

Do the following steps to install the test hookup and set the test equipment
controls:

1. Use a 50

W SMA coaxial cable and a SMA(Fe)-BNC(Fe) adapter to connect

the AWG610 Arbitrary Waveform Generator CH1 output connector to the
oscilloscope CH1 input connector (see Figure B–20).

ББ

ББ

OscilloscopeĂ(TDS700)

50 W SMAĂcoaxial cable

AWG610 Arbitrary Waveform Generator

SMA(Fe)ĆBNC(Ma)

adapter

Figure B-20: Event input and enhanced mode initial test hookup

2. Connect the ground closure to the EVENT IN connector on the AWG610

Arbitrary Waveform Generator rear panel.

3. Set the oscilloscope controls as follows:

Vertic al . . . . . . . . . . . . . . . . . . . . . . . . . CH1

CH1 coupling . . . . . . . . . . . . . . . . DC
CH1 sc ale . . . . . . . . . . . . . . . . . . 0.2 V/div
CH1 input impedance . . . . . . . . . . 50 W

Check Event Input with

Strobe Off

Advertising