Appendix B: Performance Verification

AWG610 Arbitrary Waveform Generator User Manual

B-19

BNC T

Adapter

AWG610 Arbitrary Waveform

Generator rear panel

Output

CH2 input

CH1 input

CH1 output

OscilloscopeĂ(TDS700)

SMA(Fe)ĆBNC(Ma)

adapter

50 W SMAĂcoaxial cable

50 W BNCĂcoaxial cable

50 W BNCĂcoaxial cable

Function Generator (AFG310)

Figure B-6: Triggered mode initial test hookup

4. Set the oscilloscope controls as follows:

Vertical . . . . . . . . . . . . . . . . . . . . . . . . . CH1 and CH2

CH1 coupling . . . . . . . . . . . . . . . . DC
CH1 scale . . . . . . . . . . . . . . . . . . 0.5 V/div
CH2 scale . . . . . . . . . . . . . . . . . . 2 V/div
CH1 input impedance . . . . . . . . . . 50 W
CH2 input impedance . . . . . . . . . . 1 MW

Horizontal

Sweep . . . . . . . . . . . . . . . . . . . . . 200 ns/div

Trigger

Source . . . . . . . . . . . . . . . . . . . . . CH1
Coupling . . . . . . . . . . . . . . . . . . . . DC
Slope . . . . . . . . . . . . . . . . . . . . . . Positive
Level . . . . . . . . . . . . . . . . . . . . . . +100 mV
Mode . . . . . . . . . . . . . . . . . . . . . . NORMAL