Appendix B: Performance Verification

AWG610 Arbitrary Waveform Generator User Manual

B-21

Waveform

output CH1

Trigger

signal

CH2

Figure B-7: Relationship between trigger signal and waveform output

Equipment

required

Two 50ĂW BNCĂcoaxial cables, aĂ50ĂW SMAĂcoaxial cable,

ĂaĂSMA(Fe)ĆBNC(Ma)Ăadapter,ĂaĂBNCĆTĂ(male to 2 females) adapter, a

function generator, and an oscilloscopeĂ(TDS700).

Prerequisites

The AWG610 Arbitrary Waveform Generator must meet the

prerequisites listed on page B-8.

1. Set the oscilloscope controls as follows:

Vertic al . . . . . . . . . . . . . . . . . . . . . . . . . CH1 and CH2

CH1 and CH2 coupling . . . . . . . . . DC
CH1 sc ale . . . . . . . . . . . . . . . . . . 0.5 V/div
CH2 sc ale . . . . . . . . . . . . . . . . . . 2 V/div
CH1 input impedance . . . . . . . . . . 50 W
CH2 input impedance . . . . . . . . . . 1 MW

Horizontal

Sweep . . . . . . . . . . . . . . . . . . . . . 20 ms/div

Trigger

Source . . . . . . . . . . . . . . . . . . . . . CH1
Coupling . . . . . . . . . . . . . . . . . . . . AC
Slope . . . . . . . . . . . . . . . . . . . . . . Positive
Level . . . . . . . . . . . . . . . . . . . . . . 0 V
Mode . . . . . . . . . . . . . . . . . . . . . . Auto

2. Set the function generator (AFG310) controls as follows:

Check Gated Mode