Appendix B: Performance Verification

B-22

AWG610 Arbitrary Waveform Generator User Manual

Function . . . . . . . . . . . . . . . . . . . . . . . . Square
Mode . . . . . . . . . . . . . . . . . . . . . . . . . . Continuous
Parameters . . . . . . . . . . . . . . . . . . . . . .

Frequency . . . . . . . . . . . . . . . . . . 10 kHz
Amplitude . . . . . . . . . . . . . . . . . . . 2.0 V into 50 W (4.0 V into 1 MW)
Offset . . . . . . . . . . . . . . . . . . . . . . 1.0 V into 50 W (2.0 V into 1 MW)

Output . . . . . . . . . . . . . . . . . . . . . . . . . Off

3. Follow the substeps below to set the AWG610 Arbitrary Waveform

Generator controls and to select the waveform file:

a. Push UTILITY (front-panel)

!System (bottom)!Factory Reset

(side)

!OK (side).

b. Push SETUP (front-panel)

!Run Mode

(bottom)

!Gated (side).

c.

Load the MODE.WFM file.

Refer to Loading Files on page B–10 for file loading procedures.

d. Push the RUN and CH1 output buttons.

The LEDs above the RUN button and CH1 output connector are on.

e.

Push HORIZONTAL MENU (front-panel)

! Clock (side).

f.

Push 1, 0, 0 and M (SHIFT+7) keys in this order or turn the general
purpose knob to set the internal clock frequency to 100 MHz.

g. Push the RUN and CH1 output buttons.

The LEDs above the RUN button and CH1 output connector are on.

h. Push the FORCE TRIGGER button.

Verify that the oscilloscope displays a sine wave when the FORCE TRIG-
GER button is pushed and that the output stops when the Force Trigger
button is released.

4. Follow the substeps below to check the gated mode with the gate signal:

a. Set the oscilloscope trigger source to CH2.

b. Turn on the function generator output.

c.

Verify that the oscilloscope displays a sine wave while the function
generator gate signal amplitude is High level. See Figure B–8.