Appendix B: Performance Verification

B-60

AWG610 Arbitrary Waveform Generator User Manual

f.

Verify that the reading for the high level on the oscilloscope screen is
within the range between 2.75 V and 3.25 V.

9. Do the following substeps to change the connection to the MARKER1

output:

a. Disconnect the 50

WĂSMA terminator from the MARKER1 connector.

b. Disconnect the 50

WĂSMA coaxial cable from MARKER1 connec-

tor and connect it to MARKER1.

c.

Connect the 50

WĂSMA terminator to the MARKER1 connector.

10. Repeat step 8 to check the MARKER1 output.

11. Repeat steps 1 through step 10, as replacing the MARKER1 and MARKER1

with the MARKER2 and MARKER2, to check the MARKER2 and MARK-
ER2 outputs.

12. Disconnect the oscilloscope.