The Graphical Waveform Editor

AWG610 Arbitrary Waveform Generator User Manual

3-75

Figure 3-12: Settings dialog box

Window and General are two types of editor setup parameters. Window
parameters only affects the active edit window. General parameters influence all
windows currently opened and that will be opened, whether they are active or
not. Table 3–21 describes the Window setup parameters, and Table 3–22
describes the general setup parameters.

Table 3-21: Setup window parameters

Parameters

Descriptions

Total Points

Specifies the data length of the waveform in the current window. The default

is 1000 points. The range of data points is from 512 to 8100032 and must

be a multiple of 8. If you specify a value larger than the current data length,

one or more zeros are added at the end of the data. If you specify a value

less than the current length, all data after the end data point is deleted.
The displayed value reflects data point changes resulting from any edit

operations (such as cut or paste) that increase or decrease the number of

data points in the record.

Clock

Specifies the clockfrequency used to calculate the pointĆtoĆpoint time

interval between each data point. The default setting is 100 MS/s. Note that

this clockdoes not define the waveform output frequency.

View

Selects either the Graphic or Table waveform data display mode. The

default setting is Graphic.

Table Type

Specifies to display tabular waveform data in binary or hexadecimal format.

This selection is available only when the View parameter is set to Table. The

Editor displays all data values in real numbers.