Operating Basics

2-38

AWG610 Arbitrary Waveform Generator User Manual

Table 2-14: Setup output parameter operations (cont.)

Bottom

button

Description

Side

button

Horizontal

Clock
Clock Ref

Marker 1

Delay
Marker 2

Delay

Sets the clock sample rate from 50 kS/s to 2.6 GS/s.
Sets the reference clock source to either Internal or External. A

valid external clock signal is 10 MHz

±

0.1 MHz with avoltage

level of 0.2 to 3.0 V

pĆp

.

Sets the Marker 1 signal delay value from 0 s to 1.5 ns.

Sets the Marker 2 signal delay value from 0 s to 1.5 ns.

Run Mode

Continuous

Triggered

Gated

Enhanced

Displays the Run Mode side menu for setting the instrument run

mode. Refer to the Run Mode Menu section on page 3-37 for an

explanation of the different run modes.

Trigger

Source

Slope

(or Polarity)
Level

Impedance

Interval

Sets trigger source to Internal or External. If External selected, all

other side menu items are not selectable except Interval.
Sets the trigger slope or gate polarity to Positive or Negative.

Sets the trigger signal level. The trigger level range is

±

5.0 V in

0.1 V increments.
Sets the external trigger input line impedance to either 50 W or

1 kW.
Sets trigger interval from 1.0 ms to 10.0 s.

Save/ReĆ

store

Save Setup
Resore SetĆ

up

Save the setup parameters set by SETUP window and FG mode

window as a setup file.
Restore asetup file.

To output a loaded waveform, push the CH 1 OUT and/or CH 1 OUT front-
panel button(s), then the RUN front-panel button. The LEDs near each button
light up to indicate they are enabled. The instrument outputs the waveform
depending on the Run mode. You can turn either or both channel outputs on or
off while the instrument is running by pushing the CH 1 OUT or CH 1 OUT
buttons. To stop the waveform output, push the RUN button so that the LED
turns off.

Outputting a Waveform