Marker output tests – Tektronix AWG610 User Manual

Page 418

Advertising
background image

Appendix B: Performance Verification

B-58

AWG610 Arbitrary Waveform Generator User Manual

Marker Output Tests

These procedures check the accuracy of the AWG610 Arbitrary Waveform
Generator marker output level.

Equipment

required

A 50ĂW SMA coaxial cable, anĂSMA(Fe)ĆBNC(Ma) adaptĆ

er,ĂaĂ50ĂWĂSMAĂterminator,Ăand an oscilloscope.

Prerequisites

The

AWG610 Arbitrary Waveform Generator

must meet the

prerequisites listed on page B-8.

NOTE. Connect a 50W SMA terminator to the inverted marker output connector
during the marker output tests.

Do the following steps to install the test hookup and set the test equipment
controls:

1. Use an SMA coaxial cable and an SMA-to-BNC adapter to connect the

AWG610 Arbitrary Waveform Generator MARKER1 connector to the
oscilloscope CH1 input connector (see Figure B–31).

2. Connect a 50

WĂSMA terminator to the MARKER1 connector.

ББ

ББ

AWG610 Arbitrary Waveform Generator

50 W SMA coaxial cable

50 WĂSMA terminator

OscilloscopeĂ(TDS700)

SMA(Fe)ĆBNC(Ma)

adapter

Figure B-31: Marker output initial test hookup

3. Set the oscilloscope controls as follows:

Vertical . . . . . . . . . . . . . . . . . . . . . . . . . CH1

CH1 coupling . . . . . . . . . . . . . . . . DC
CH1 scale. . . . . . . . . . . . . . . . . . 1 V/div
CH1 input impedance . . . . . . . . . . 50 W
CH1 offse t . . . . . . . . . . . . . . . . . . 0 V