10 mhz reference output levelătests – Tektronix AWG610 User Manual

Page 416

Advertising
background image

Appendix B: Performance Verification

B-56

AWG610 Arbitrary Waveform Generator User Manual

10 MHz Reference Output LevelĂTests

These procedures check the AWG610 Arbitrary Waveform Generator clock
output signal.

Equipment

required

A 50ĂW BNC coaxial cable and an oscilloscope.

Prerequisites

The instrument must meet the prerequisites listed on page B-8.

Do the following steps to install the test hookup and set the test equipment
controls:

1. Use the 50

W

BNC coaxial cable to connect the AWG610 Arbitrary

Waveform Generator 10 MHz REF OUT output to the oscilloscope input
connector (see Figure B–29).

10 MHz REF OUT

AWG610 Arbitrary Waveform

Generator rear panel

OscilloscopeĂ(TDS700)

50 W BNCĂcoaxial cable

Figure B-30: 10MHz Reference output level initial test hookup

2. Set the oscilloscope controls as follows:

Vertical . . . . . . . . . . . . . . . . . . . . . . . . . CH1

CH1 coupling . . . . . . . . . . . . . . . . DC
CH1 scale . . . . . . . . . . . . . . . . . . 0.2 V/div
CH1 offset . . . . . . . . . . . . . . . . . . 0 V
CH1 input impedance . . . . . . . . . . 50 W

Horizontal

Sweep . . . . . . . . . . . . . . . . . . . . . 25 ns/div

Trigger

Source . . . . . . . . . . . . . . . . . . . . . CH1
Coupling . . . . . . . . . . . . . . . . . . . . AC
Slope . . . . . . . . . . . . . . . . . . . . . . Positive
Level . . . . . . . . . . . . . . . . . . . . . . 0 V
Mode . . . . . . . . . . . . . . . . . . . . . . Auto

Advertising