Appendix B: Performance Verification

AWG610 Arbitrary Waveform Generator User Manual

B-41

3. Set the oscilloscope controls as follows:

Push the Default Setup (front).

Verti cal . . . . . . . . . . . . . . . . . . . . . . . . . CH1 and CH2

CH1 coupling . . . . . . . . . . . . . . . . DC
CH1 scale . . . . . . . . . . . . . . . . . . 500 mV/div
CH1 input impedance . . . . . . . . . . 50 W
CH2 scale . . . . . . . . . . . . . . . . . . 2 V/div
CH2 input impedance . . . . . . . . . . 1 MW

Horizontal

Sweep . . . . . . . . . . . . . . . . . . . . . 2 ms/div

Trigger

Source . . . . . . . . . . . . . . . . . . . . . CH1
Coupling . . . . . . . . . . . . . . . . . . . . DC
Slope . . . . . . . . . . . . . . . . . . . . . . Positive
Level . . . . . . . . . . . . . . . . . . . . . . +100 mV
Mode . . . . . . . . . . . . . . . . . . . . . . Auto

4. Set the voltage source controls as follows:

Function . . . . . . . . . . . . . . . . . . . . . . . . Pulse
Mode . . . . . . . . . . . . . . . . . . . . . . . . . . Continuous
Parameter

Frequency . . . . . . . . . . . . . . . . . . 100 Hz
Amplitude . . . . . . . . . . . . . . . . . . . 1.0 V (2 V i n open circuit)
Offset . . . . . . . . . . . . . . . . . . . . . (Adjust such as a pulse 4.65 V in amplitude

referenced to ground)

Output . . . . . . . . . . . . . . . . . . . . . . . . . Off

5. Set the AWG610 Arbitrary Waveform Generator and load the waveform file.

a. Push UTILITY (front-panel)

!System (bottom)!Factory Reset

(side)

!OK (side).

b. Push SETUP (front-panel)

!Run Mode (bottom)!Triggered (side).

c.

Load the TRIG.WFM file.

Refer to Loading Files on page B–10 for file loading procedures.

6. Push the RUN and CH1 OUT buttons.

The LEDs above the RUN button and CH1 output connector are on.

Advertising