Agilent Technologies E1300B User Manual

Page 134

Advertising
background image

:BOOT:COLD

DIAGnostic:BOOT:COLD causes the System Instrument to restart (re-boot).
Configurations stored in non-volatile memory and RS-232 configurations are
reset to their default states:

DRAM, NRAM, and RDISk memory segments are cleared

Serial Interface parameters set to:

BAUD 9600
BITS 8
PARity NONE
SBITs 1
DTR ON
RTS ON
PACE XON

Serial 0 Owner = system

NOTE

Resetting the serial interface parameters takes about 0.01 seconds for the
built-in serial port and 0.75 seconds per serial plug-in card. While this is taking
place the System Instrument will still respond to serial polls. If you are using a
serial poll to determine when the cold boot cycle is complete, you should insert a
delay of 1 second per plug-in serial card (E1324) before polling the system
instrument. This will prevent incorrectly determining that the system instrument
has completed its boot cycle.

Comments

The System Instrument goes through its power-up self tests.

Related Commands: DIAG:BOOT:WARM

Example

Re-booting the System Instrument (cold)

DIAG:BOOT:COLD

force boot

DIAGnostic:BOOT:COLD

7-6 System Instrument Command Reference