Error Messages and Causes

Code

Message

Cause

2110

Config error 10, Insufficient
system memory

Too many instruments installed for the amount
of RAM installed in the mainframe. Cannot
configure instruments. O nly the system
instrument is started.

2111

Config error 11, Invalid
instrument address

A device’s logical address is not a multiple of 8
and the device is not part of a combined
instrument.

2113

Config error 13, Logical
address or IACK switch set
wrong

D uplicate logical addresses set or interrupt
bypass switches set improperly. O nly the system
instrument is started.

2129

Config warning, Sysfail
detected

A device was asserting SYSFAIL on the
backplane during startup.

2130

Config error 30, Pseudo
instrument logical address
unavailable

A physical device has the same logical address as
IBASIC (240)

2131

Config error 32, File system
start up failed

Insufficient system resources to allow the
IBASIC file system to start.

2145

Config warning, Non-volatile
RAM contents lost

NVRAM was corrupted or a cold boot was
executed.

2148

Config warning, D river RAM
contents lost

D river RAM was corrupted or a cold boot was
executed.

2202

U nexpected interrupt from
non-message based card

A register based card interrupted when an
interrupt service routine had not been set up.

2809

Interrupt line has not been
set up

A D IAG :INT:ACT or D IAG :INT:RESP
command was executed before setting the
interrupt with D IAG :INT:SET.

B-4 Error Messages