:COMMunicate

:SERial[n] :CONTrol

:RTS?

SYSTem:COMMunicate:SERial[ n] :CONTrol:RTS? returns the current setting
for RTS line control.

Example

Checking the setting of RTS control.

SYST:COMM:SER0:CONT:RTS?
enter statement

statement enters the string
"ON", "OFF", "STAN", or "IBF"

:COMMunicate

:SERial[n] [:RECeive]

:BAUD

SYSTem:COMMunicate:SERial[ n] [ :RECeive] :BAUD < baud_rate> Sets the
baud rate for the serial port.

Parameters

Parameter

Name

Parameter

Type

Range of

Values

Default

Units

baud

numeric

300 | 1200 | 2400 | 4800 |

9600 | 19200 | MIN | MAX

none

Comments

Attempting to set baud to other than those values shown will result in an
error -222.

DIAG:BOOT:COLD will set

BAUD to 9600.

*RST condition: No change.

Example

Setting the baud rate to 1200.

SYST:COMM:SER0:BAUD 1200

:COMMunicate

:SERial[n] [:RECeive]

:BAUD?

SYSTem:COMMunicate:SERial[ n] [ :RECeive] :BAUD? [ MIN | MAX] returns:

The current baud rate setting if no parameter is sent.

The maximum allowable setting if MAX is sent.

The minimum allowable setting if MIN is sent.

Example

Querying the current baud rate.

SYST:COMM:SER0:BAUD?
enter statement

statement enters a numeric
value

SYSTem :COMMunicate :SERial[n] :CONTrol :RTS?

System Instrument Command Reference 7-39