:COMMunicate

:SERial[n] [:RECeive]

:PACE :THReshold

:STARt

SYSTem:COMMunicate:SERial[ n] [ :RECeive] :PACE:THReshold:STARt
< char_count
> configures the input buffer level at which the specified interface
may send the XON character (ASCII 11

16

), assert the DTR line, and/or assert

the RTS line.

Parameters

Parameter

Name

Parameter

Type

Range of

Values

Default

Units

char_count

numeric

1 through 99 for built-in

1 through 8191 for E1324A

none

Comments

To determine the size of the input buffer of the serial interface you are
using, send SYST:COMM:SER[n]:PACE:THR:START? MAX. The
returned value will be the buffer size less one.

• …

STARt must be set to less than

STOP.

The

THR:STAR command has no effect unless

PACE:PROT XON,

CONT:DTR IBF, or

CONT:DTR IBF has

been sent.

Related Commands:

PACE:PROT XON | NONE,

CONT:DTR,

CONT:RTS

*RST Condition: No change

Example

Set interface to send XON when input buffer contains 10 characters.

SYST:COMM:SER0:PACE:PROT XON

SYST:COMM:SER0:PACE:THR:STAR 10

:COMMunicate

:SERial[n] [:RECeive]

:PACE :THReshold

:STARt?

SYSTem:COMMunicate:SERial[ n] [ :RECeive] :PACE:THReshold:STARt?
[ MIN | MAX]
returns:

The current start threshold if no parameter is sent.

The maximum allowable setting if MAX is sent.

The minimum allowable setting if MIN is sent.

Comments

To determine the size of the input buffer of the serial interface you are
using, send SYST:COMM:SER[n]:PACE:THR:START? MAX. The
returned value will be the buffer size.

Example

Return current start threshold

SYST:COMM:SER0:PACE:THR:STAR?

query for threshold value

enter statement

statement enters a numeric
value

SYSTem:COMMunicate :SERial[n] [:RECeive] :PACE :THReshold :STARt

7-42 System Instrument Command Reference