Appendix

B

Error Messages

Using This Appendix

This appendix shows how to read an instrument’s error queue, discusses the
types of command language-related error messages, and provides a table of all
of the System Instrument’s error messages and their probable causes.

Reading an Instrument’s Error Queue . . . . . . . . . . . . . . . . . . . . B-1

Error Types . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-2

Start-up Error Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-5

1

Reading an
Instrument’s Error
Queue

Executing the SYST:ERR? command reads the oldest error message from the
instrument’s error queue and erases that error from the error queue. The
SYST:ERR? command returns response data in the form:

< error number> ,"< error description string> ".

Example error message; -113,"Undefined header"

Positive error numbers are specific to an instrument. Negative error numbers
are command language-related and discussed in the next section "Error
Messages". Command language-related errors also set a corresponding bit in the
Standard Event Status Register (refer to "Instrument Status" in Chapter 4 for
more information).

Example: Reading the Error

Queue

This program reads all errors (one error at a time, oldest to newest) from the
System Instrument’s error queue. After reading each error, that error is
automatically erased from the queue. When the error queue is empty, this
program returns:

+ 0,"No error"

.

10 OPTION BASE 1

20 DIM Message$[256]

Create array for error message

30 REPEAT

Repeat next 3 lines until error
number = 0

40 OUTPUT 70900;"SYST:ERR?"

Read error number & message

50 ENTER 70900;Code,Message$

Enter error number & message

60 PRINT Code,Message$

Print error number & message

70 UNTIL Code= 0
80 END