:COMMunicate

:SERial[n] [:RECeive]

:BITS

SYSTem:COMMunicate:SERial[ n] [ :RECeive] :BITS < bits> Sets the number
of bits to be used to transmit and receive data.

Parameters

Parameter

Name

Parameter

Type

Range of

Values

Default

Units

bits

numeric

7| 8| MIN| MAX

none

Comments

Attempting to set bits to other than those values shown will result in an
error -222.

While this command operates independently of either the

PARity:TYPE or

SBITs commands, there are two combinations

which are disallowed because of their data frame bit width. The following
table shows the possible combinations:

BITS

PARity:TYPE

SBITs

Frame Bits

7

NONE

1

9 - disallowed

7

NONE

2

10

7

Yes

1

10

7

Yes

2

11

8

NONE

1

10

8

NONE

2

11

8

Yes

1

11

8

Yes

2

12 - disallowed

DIAG:BOOT:COLD will set

BITS to 8.

Related Commands: SYST:COMM:SER[n]:PARity

*RST Condition: No change

Example

Configuring data width to 7 bits.

SYST:COMM:SER0:BITS 7

:COMMunicate

:SERial[n] [:RECeive]

:BITS?

SYSTem:COMMunicate:SERial[ n] [ :RECeive] :BITS? [ MIN | MAX] returns:

The current data width if no parameter is sent.

The maximum allowable setting if MAX is sent.

The minimum allowable setting if MIN is sent.

Example

Querying the current data width.

SYST:COMM:SER0:BITS?

enter statement

statement enters 7 or 8

SYSTem:COMMunicate :SERial[n] [:RECeive] :BITS