:WRITe

VXI:WRITe < logical_addr> ,< register_addr> ,< data> allows access to the
entire 64 byte A16 register address space for the device specified by
logical_addr. Since the VXIbus system is byte-addressed, while the registers are
16 bits wide, registers are specified by even addresses only. This method of
identifying registers follows the VXIbus standard format.

Parameters

Parameter

Name

Parameter

Type

Range of

Values

Default

Units

logical_addr

decimal

numeric

Must round to 0 through 255

none

register_addr

numeric

must round to an even value

from 0 through 62 (3E

h

)

none

data

numeric

must round to -32768 to 32767

(0 to FFFF

h

)

none

Comments

Specifying an odd register address will cause an error 2003,"Invalid word
address".

Specifying a logical address not currently in use in the system will cause
an error 2005,"No card at logical address".

Logical_addr must be specified in decimal. Register_addr and data may
be specified in decimal, hex (# H), octal (# Q), or binary (# B).

This command has been retained for compatibility with existing
programs. For new programs you should use the VXI:REG:WRIT
command.

Accesses are 16-bit non-privileged data accesses.

Related Commands: VXI:READ?, VXI:REG:WRIT

Example

Write a value into a device’s device dependent register.

VXI:WRIT 8,24,# H4200

write hex 4200 (16,896
decimal) to register 24 of device
at Logical Address 8

VXI:WRITe

7-64 System Instrument Command Reference