Agilent Technologies E1300B User Manual

Page 148

Advertising
background image

ON or 1 specify that interrupt handling is to be set up for the specified
interrupt line. OFF or 0 indicate that interrupt handling of the specified
line is to be done by the operating system.

Related Commands: DIAG:INT:ACT, DIAG:INT:PRIority[n],
DIAG:INT:RESP?

*RST Condition: DIAG:INT:SETup[n] OFF (for all lines)

Example

Setup and wait for VXI interrupt response on line 2.

DIAG:INT:PRI2 5

set priority to 5 on line 2

DIAG:INT:SETUP2 ON

handle interrupt on line 2

.

code which will

.

initiate an action

.

resulting in an interrupt

DIAG:INT:RESP?

Read the acknowledge response

:INTerrupt:SETup[n]?

DIAGnostic:INTerrupt:SETup[ n] ? Returns the current state set by
DIAG:INT:SETUP[n] < mode> , for the VXI interrupt line specified by [n] in

SETup[n]?.

Comments

• …

SETup1? through

SETup7? specify the VXI interrupt lines 1

through 7.

Sending SETup? without an [n] value specifies VXI interrupt line 1.

If 1 is returned, interrupt handling is set up for the specified interrupt
line using the System Instrument (DIAGnostic:INTerrupt commands). If
0 is returned, interrupt handling is done by the operating system.

Related Commands: DIAG:INT:SETup[n], DIAG:INT:PRIority[n],
DIAG:INT:ACT, DIAG:INT:RESP?

Example

Determine interrupt setup for line 4.

DIAG:INT:SETUP4?
enter statement

statement returns 0 or 1

DIAGnostic:INTerrupt:SETup[n]?

7-20 System Instrument Command Reference

Advertising