Error Messages and Causes

Code

Message

Cause

-101

Invalid character

U nrecognized character in specified parameter.

- 102

Syntax error

Command is missing a space or comma between
parameters

- 103

Invalid separator

Command parameter is separated by some
character other than a comma.

- 104

D ata type error

The wrong data type (i.e. number, character,
string expression) was used when specifying a
parameter.

- 108

Parameter not allowed

Parameter specified in a command which does
not require one.

- 109

Missing parameter

No parameter specified in the command in
which a parameter is required.

- 113

U ndefined header

Command header was incorrectly specified.

- 123

Numeric overflow

A parameter specifies a value greater than the
command allows.

- 128

Numeric data not allowed

A number was specified for a parameter when a
letter is required.

- 131

Invalid suffix

Parameter suffix incorrectly specified
(e.g. .5SECO ND rather than .5S or .5SEC).

- 138

Suffix not allowed

Parameter suffix is specified when one is not
allowed.

- 141

Invalid character data

The discrete parameter specified is not allowed
(e.g. TRIG :SO U R INT - INT is not a choice.)

- 178

Expression data not allowed

A parameter other than the channel list is
enclosed in parentheses.

- 211

Trigger ignored

Trigger occurred while the Pacer is in the idle
state, or a trigger occurred from a source other
than the specified source.

- 222

D ata out of range

The parameter value specified is too large or too
small.

- 224

Illegal parameter value

The numeric value specified is not allowed.

- 240

H ardware error

H ardware error detected during power-on cycle.
Return multimeter to Agilent for repair.

- 310

System error

If caused by *D MC, then macro memory is full.

- 350

Too many errors

The error queue is full as more than 30 errors
have occured.

- 410

Q uery interrupted

D ata is not read from the output buffer before
another command is executed.

- 420

Q uery unterminated

Command which generates data not able to
finish executing due to a multimeter
configuration erorr.

- 430

Q uery deadlocked

Command execution cannot continue since the
mainframe’s command input, and data output
buffers are full. Clearing the instrument restores
control.

1500

External trigger source
already allocated

"Event In" signal already allocated to another
instrument such as a Switchbox.

2002

Invalid logical address

A value less than 0 or greater than 255 was
specified for logical address.

2003

Invalid word address

An odd address was specified for a 16 bit read or
write. Always use even addresses for 16 bit
(word) accesses.

2005

No card at logical address

A non-existent logical address was specified with
the VXI:READ ? or VXI:WRITE command.

2101

Failed D evice

VXI device failed its self test.

2102

U nable to combine device

D evice type can not be combined into an
instrument such as a scanning voltmeter or a
switchbox.

2103

Config warning, D evice
driver not found

ID of device does not match list of drivers
available. Warning only.

2105

Config error 5, A24 memory
overflow

More A24 memory installed in the mainframe
than can be configured into the available A24
memory space.

2108

Config error 8, Inaccessible
A24 memory

A24 memory device overlaps memory space
reserved by the mainframe’s operating system.

Table B-2. Error Messages and Causes

Error Messages B-3