Agilent Technologies E1300B User Manual

Page 142

Advertising
background image

Example

Loading Dynamic Configuration information into an allocated RAM segment.

DIAG:NRAM:CRE 6

Allocate a segment of user
RAM

DIAG:BOOT:WARM

Re-boot system to complete
allocation

DIAG:NRAM:ADDR?

query starting address

enter value to variable X

get starting address into X

DIAG:DOWN < value of X> ,table data

download table data

VXI:CONF:DCTAB < value of X>

link configuration table to
configuration algorithm

DIAG:BOOT:WARM

Re-boot to set new
configuration

:DOWNload:SADDress

DIAGnostic:DOWNload:SADDress < address> ,< data> writes data to
non-volatile User RAM at a single address specified by address, and writes data
to devices with registers in A16 address space.

Parameters

Parameter

Name

Parameter

Type

Range of

Values

Default

Units

address

numeric

0 to 16,777,215 (# HFFFFFE)

none

data

arbitrary block

program data

See "Parameter Types", in the

beginning of this chapter

none

Comments

Most computers terminate an OUTPUT, PRINT, or WRITE statement
with a carriage return or carriage return and line feed. These
End-Of-Line characters must be accounted for or suppressed using an
appropriate IMAGE or FORMAT statement. Some helpful methods:

Control the End-Of-Line characters with format statements.
Use the Definite Length Arbitrary Block Program Data format to send

your data rather than the Indefinite Length Arbitrary Block Program
Data
format.

A register address in A16 address space can be determined by:

1FC000

16

+ (LADDR * 64) + register_number

where 1FC000

16

is the base address in the System Instrument A16

address space, LADDR is the device logical address, 64 is the number of
address bytes per device, and register_number is the register to which the
data is written.

If the device is an A24 device, the address can be determined using the
VXI:CONF:DLISt command to find the base address in A24, and then
adding the register_number to that value. A24 memory between address
200000

16

and address E00000

16

is directly addressable by the Controller.

Address may be specified in decimal, hex (# H), octal (# Q), or binary
(# B) formats. DOWNload is done by word (16 bit) access so address
must be even.

Related Commands: DIAG:UPLoad:SADDress?

DIAGnostic:DOWNload:SADDress

7-14 System Instrument Command Reference

Advertising