Agilent Technologies E1300B User Manual

Page 181

Advertising
background image

Comments

The following table explains the possible choices.

Parameter Value

Source of Trigger

BUS

Group Execute Trigger (GET) bus command,
*TRG common command, or TRIGger
command.

EXTernal

“Event In” signal at rear panel BNC
connector, or TRIGger command.

HOLD

Only the TRIGger command will cause
trigger.

IMMediate

The trigger signal is always true (continuous
triggering).

While an instrument which uses the "Event In" signal has EXT set, no
other instrument which uses the "Event In" signal may set EXT, or an
error 1500 "External trigger source already allocated" will result.

While TRIG:SOUR is IMM, you need only INITiate the trigger system to
start the Pacer.

Related Commands: ABORt, INITiate, *TRG

*RST Condition: TRIG:SOUR IMM

Example

Specifying the Trigger Source.

TRIG:SOUR HOLD

trigger source is TRIG
command

SOUR:PULS:COUN 1E3

output 1000 Pacer pulses

SOUR:PULS:PER .1 S

pulse period set to .1 second

TRIG:DELAY .75 S

start Pacer .75 sec after trigger

INIT

go to Wait For Trigger state

TRIG

trigger the Pacer to output
pulses.

:SOURce?

TRIGger:SOURce? returns the current trigger source configuration. Response
data can be one of; BUS, EXT, HOLD, or IMM. See the TRIG:SOUR
command for more response data information.

Example

Querying the Trigger Source.

TRIG:SOUR HOLD

trigger source is TRIG
command

TRIG:SOUR?

ask System Instrument to
return trigger source
configuration

enter statement

input selection of trigger source

TRIGger :SOURce?

System Instrument Command Reference 7-53