[SOURce]

The System Instrument contains a Pacer which produces TTL level pulses. The
SOURCE command subsystem controls the number and period of these pulses.
The output of the Pacer is available at the rear-panel BNC connector labeled
“Pacer Out”.

Subsystem Syntax

[SOURce]

:PULSe

:COUNt < count>
:COUNt? [MIN | MAX]
:PERiod < period>
:PERiod? [MIN | MAX]

:PULSe:COUNt

SOURce:PULSe:COUNt < count> sets the number of Pacer pulses that are
generated when the trigger condition is satisfied.

Parameters

Parameter

Name

Parameter

Type

Range of

Values

Default

Units

count

numeric

1 to 8,388,607| 9.9E37|

INFinity| MIN| MAX

none

Comments

When count is set to INFinity or 9.9E37, pulses are continuous.

Related Commands: ABORT, INIT, TRIG

*RST Condition: SOUR:COUN 1

Example

Setting the Pacer pulse count.

TRIG:SOUR HOLD

trigger source is TRIG
command

SOUR:PULS:COUN 1E3

output 1000 Pacer pulses

SOUR:PULS:PER .1 S

pulse period set to .1 second

INIT

go to Wait For Trigger state

TRIG

trigger the Pacer to output
pulses

:PULSe:COUNt?

SOURce:PULSe:COUNt? [ MIN | MAX] returns:

The current count if no parameter is sent.

The maximum allowable count if MAX is sent.

The minimum allowable count if MIN is sent.

Example

Querying the pulse count.

SOUR:PULS:COUN 1E3

output 1000 Pacer pulses

SOUR:PULS:COUN?

query system for pulse count

retrieve value

[SOURce]:PULSe:COUNt

7-30 System Instrument Command Reference