Agilent Technologies E1300B User Manual

Page 93

Advertising
background image

multimeter to an internal multiplexer. Connections are shown in Figure 4-4. This
method synchronizes the computer to the instruments and relies on the
computer to enter each reading and advance to the next channel in the scan list.
The sequence of operation is:

1. INIT (line 50) closes channel number 100.
2. The channel closure causes a pulse on Trig Out which triggers the

multimeter to take a reading.

3. When the reading is complete it is sent to the computer (lines 60 to 80).
4. The computer sends Group Execute Trigger to the multiplexer (line 90);

this advances to the next channel in the scan list.

5. Steps 2 through 4 are repeated until all channels have been scanned and

readings taken.

10 OUTPUT 722;"TRIG EXT;DCV"

Set multimeter to external trigger, DC voltage measurements

20 OUTPUT 70914;"OUTP ON"

Enable Trig Out port

30 OUTPUT 70914;"TRIG:SOUR BUS"

Set multiplexer to advance scan on Group Execute Trigger or
*TRG

40 OUTPUT 70914;"SCAN (@100:115)"

Specify scan list (channels 100
to 115)

50 OUTPUT 70914;"INIT"

Close first channel (starts
scanning cycle)

60 FOR I= 1 TO 16

Loop through following lines
16 times

70 ENTER 722;A

Enter reading (computer waits
until reading taken & received)

80 PRINT A

Print reading

90 TRIGGER 70914

Trigger multiplexer; advances
to next channel

100 NEXT I

110 END

Using the Mainframe 4-5

Advertising