Operator’s Manual

922131-00 Rev A

41

Cursor Measurement Selections

Time cursors are vertical lines that you move horizontally to measure
the difference in time or frequency values between the cursors.

Voltage cursors appear as horizontal lines on the display and measure
the vertical parameters.

Measurement values are shown in the upper-left corner of the grid
display and are explained as follows:

Shown in Manual Cursor Mode
▲V - the vertical space between Cursor A and Cursor B (Voltage value
between the two cursors)

CurB - the horizontal position of Cursor B (in Volts)

CurA - the horizontal position of Cursor A (in Volts)

Shown in Track Cursor Mode

A→T - the horizontal position of Cursor A (Time cursor centered around
the midpoint of the screen)

A→V - the vertical position of Cursor A (Voltage cursor centered on the
channel ground level)

B→T - the horizontal position of Cursor B (Time cursor centered on the
midpoint of the screen)

B→V - the vertical position of Cursor B (Voltage cursor centered around
the channel ground level)
▲T - Horizontal space between Cursor A and Cursor B (Time value
between the two cursors)

1/▲T - the reciprocal of the horizontal space between Cursor A and
Cursor B
▲V - the vertical space between Cursor A and Cursor B (Voltage value
between the two cursors)