Adjusting sensitivity, Operator’s manual – Teledyne LeCroy WaveAce 1000_2000 User Manual

Page 31

Advertising
background image

Operator’s Manual

922131-00 Rev A

25

Adjusting Sensitivity

You can set the sensitivity of the Volts/Div using the front panel knob. When
the sensitivity is Coarse (or fixed - the default setting), the gain adjustment is
set to 1-2-5 increments from 2 mV/div, 5 mV/div, 10 mV/div to 5 V/div.
Pushing the Volts/Div knob sets the sensitivity to Fine (variable), and changes
the resolution to small steps between the coarse settings.

NOTE: The vertical scale readout displays the actual Volts/Div setting when
Fine(variable) is selected. Changing the setting to Coarse (fixed) does not
change the vertical scale until the Volts/Div control is adjusted.

1. Set the sensitivity of the Volts/Div front panel knob by turning on the

specific channel. Press the desired channel button - CH1 or CH2 (or CH3
or 4 on 4 channel models).

2. Now, press the Volts/Div option button on page 1/3 of the channel

menu, and then select Coarse (fixed) or Fine (variable).

Advertising