Electromagnetic Emissions:

CISPR 11:2003, Radiated and Conducted Emissions Group 1, Class A

2 3

EN 61000-3-2:2006 Harmonic Current Emissions, Class A

EN 61000-3-3/A2:2005 Voltage Fluctuations and Flickers, Pst = 1

Electromagnetic Immunity:

EN 61000-4-2:2001 Electrostatic Discharge, 4 kV contact, 8 kV air, 4 kV
vertical/horizontal coupling planes

4

EN 61000-4-3:2006 RF Radiated Electromagnetic Field, 3 V/m, 80-1000 MHz; 3
V/m, 1400 MHz - 2 GHz; 1 V/m, 2 GHz - 2.7 GHz

4

EN 61000-4-4:2004 Electrical Fast Transient/Burst, 1 kV on power supply lines,
0.5 kV on I/O signal data and control lines

4

EN 61000-4-5:2006 Power line Surge, 1 kV AC Mains, L-N, L-PE, N-PE

4

EN 61000-4-6:2007 RF Conducted Electromagnetic Field, 3 Vrms, 0.15 MHz -
80 MHz

4

EN 61000-4-11:2004 Mains Dips and Interruptions, 0%/1 cycle, 70%/25 cycles,
0%/250 cycles

4 5

1 To ensure compliance with all applicable EMC standards, high quality shielded
interface cables should be used.

2 Emissions which exceed the levels required by this standard may occur when the
oscilloscope is connected to a test object.

3 This product is intended for use in nonresidential areas only. Use in residential areas
may cause electromagnetic interference.

4 Meets Performance Criteria “B” limits of the respective standard: during the
disturbance, product undergoes a temporary degradation or loss of function or
performance which is self-recoverable.

5 Performance Criteria “C” applied for 70%/25 cycle voltage dips and for 0%/250 cycle
voltage interruption test levels per EN61000-4-11.