Utility Menu - Page 2/4

D

O

S

ELF

C

AL

1. Press the Do Self Cal option button to perform a self-calibration. The Do

Self Cal screen is shown.

2. As the screen indicates, disconnect anything plugged into input

connectors (except for the power cable) before pressing the SINGLE
Front Panel (on page 7) button and performing a self-calibration. Press
the RUN/STOP front panel button to exit the Do Self Cal screen.

3. After pressing the SINGLE front panel button, a status bar shows the

self-calibration progress for each channel on the instrument.