Voltage measurement parameters – Teledyne LeCroy WaveAce 1000_2000 User Manual

Page 50

Advertising
background image

WaveAce 1000/2000

44

922131-00 Rev A

Voltage Measurement Parameters

First select a Source Channel for your Voltage measurement parameter by
pressing the Source option button. Now, use the Type option button to choose
from the following Voltage measurement parameters for your selected source.

Vpp

- Difference between highest and lowest points in the

waveform.

Vmax

- Measures highest point in waveform. Unlike top, it does

not assume the waveform has two levels.

Vmin

- Measures the lowest point in a waveform. Unlike base, it

does not assume the waveform has two levels.

Mean

- Average of the data for a time domain waveform.

Computed as centroid of distribution for a histogram.

Vrms

- Root Mean Square of data between the cursors - about

the same as sdev for a zero-mean waveform.

Crms

- Cyclic root mean square: Computes the square root of

the sum of squares of data values divided by number of points.
Contrary to rms, calculation is performed over an integer number of
cycles, eliminating bias caused by fractional intervals.

Vtop

- Higher of two most probable states, the lower being

base; it is characteristic of rectangular waveforms and represents the
higher most probable state determined from the statistical distribution
of data point values in the waveform.

Advertising