Vaisala HMT330 User Manual

Page 101

Advertising
background image

Chapter 4 _________________________________________________________________ Operation

VAISALA ________________________________________________________________________ 99

FST

To output the state of optional probe heating and chemical purge in
connection with SEND and R commands enter:

FST [x]<cr>

where

x

=

ON or OFF (default)

Examples:

>fst on
Form. status : ON

>send
N 0 RH= 40.1 %RH T= 24.0 'C Td= 9.7 'C Tdf= 9.7 'C a=
8.7 g/m3 x= 7.5
g/kg Tw= 15.6 'C ppm= 11980 pw= 12.00 hPa pws= 29.91 hPa
h= 43.2 kJ/kg

>purge
Purge started, press any key to abort.

>send
S 134 RH= 40.2 %RH T= 24.1 'C Td= 9.8 'C Tdf= 9.8 'C a=
8.8 g/m3 x= 7.5
g/kg Tw= 15.7 'C ppm= 12084 pw= 12.10 hPa pws= 30.11 hPa
h= 43.5 kJ/kg
>

where the state of the probe is indicated by the following letters and values:

N ... xxx = Normal operation where h xxx = Probe heat power
X ... xxx = Sensor heating

where xxx = Sensor temperature (ºC)

H ... xxx = Chemical purge where xxx = Sensor temperature (ºC)
S ... xxx

= Sensor cooling

after purge

where xxx = Sensor temperature (ºC)

For more information on chemical purge, see section Chemical Purge
(Optional) on pa
ge 133.