Vaisala HMT330 User Manual

Page 141

Advertising
background image

Chapter 4 _________________________________________________________________ Operation

VAISALA _______________________________________________________________________ 139

To configure the sensor heating use the XHEAT command without
parameters. Enter the values after question mark. The available ranges
include the following:

Extra heat RH -limit (heating
function starts-up above the
setpoint)

0 ... 100 %RH (default: 95 %RH)

Extra heating temperature

0 ... 200 ºC (default: 100 ºC)

Extra heating time

0 ... 255 s (default: 30 s)

Example:


>xheat
Extra heat : OFF
Extra heat RH : 95 ? 90
Extra heat temp: 100 ? 85
Extra heat time: 30 ? 10
>xheat on
Extra heat : ON
>

Advertising