Appendix A _________________________________ Probe Installation Kits and Installation Examples

VAISALA _______________________________________________________________________ 191

Example of Installation Through Roof

Outdoor
environment

Process:

For example
+85 ºC, 85 %RH

0507-015

Figure 101

Example of Installation through Roof

The following numbers refer to Figure 101:

1

=

Insulated probe cables

2

=

Sealings

3

=

Roof

4

=

Cable gland for temperature probe (for example: AGRO 1100.12.91.065)

5

=

Temperature probe

6

=

Relative humidity probe

7

=

Cable gland for relative humidity probe (for example: AGRO 1160.20.145)

8

=

Plastic adapter to protect probes from condensation water coming from the
pipe. Diameter slightly smaller than tube diameter.

9

=

Plastic tube for probe (2 pcs)

10 =

Stainless steel tube coming through the roof

11 =

Two thread bars holding the plastic adapter

12 =

Insulated pipe ending