Chapter 4 _________________________________________________________________ Operation

VAISALA _______________________________________________________________________ 105

Table 28

Conversion Factors for Pressure Units

From

To: hPa

mbar

1

Pa N/m2

0.01

mmHg torr

1.333224

inHg

33.86388

mmH

2

O

0.09806650

inH

2

O

2.490889

atm

1013.25

at

980.665

bar

1000

psia

1)

68.94757

1) psia = psi absolute.

For example, 29.9213 inHg = 29.9213 × 33.86388 = 1013.25 hPa.

Limit RH Output Range

Use the serial line command RHLIMIT if you want to strictly limit the
allowed value of the relative humidity (RH) quantity to range 0 ... 100%.
The limit is implemented on all outputs and the display. Other quantities
are not affected, only RH. Note that seeing values above 100 %RH is
useful for detecting condensation on the sensor, and possible
measurement drift in very high humidities. For this reason, Vaisala
recommends leaving this feature at the default setting (off).

RHLIMIT [ON/OFF]<cr>

where

ON = Allowed range of RH output is 0 ... 100 %.
OFF = Allowed range of RH output is -5 ... 110 % (default).

Example (view current setting):

>rhlimit ?
RH limit 0...100: OFF
>

Example (enable RH limit):

>rhlimit on
RH limit 0...100: ON
>