Technical specifications of optional modules, Power supply module, Analog output module – Vaisala HMT330 User Manual

Page 178: Relay module, Technical specifications of optional modules on pa

Advertising
background image

User's Guide _______________________________________________________________________

176 __________________________________________________________________ M210566EN-K

Technical Specifications of Optional
Modules

Power Supply Module

Operating voltage

100 ... 240 VAC 50/60 Hz

Connections

screw terminals for
0.5 ... 2.5 mm

2

wire

(AWG 20 ... 14)

Bushing

for 8 ... 11 mm diameter cable

Operating temperature

-40 ... +60 °C (-40 ... +140 °F)

Storage temperature

-40 ... +70°C (-40 ... +158 °F)

UL file number

E249387

Analog Output Module

Outputs

0 ... 20 mA, 4 ... 20 mA,
0 ... 1 V, 0 ... 5 V, 0 ... 10 V

Operating temperature range

-40 ... +60 °C (-40 ... +140 ºF)

Power consumption

U

out

0 ... 1 V

max 30 mA

U

out

0 ... 5V/0 ... 10V

max 30 mA

I

out

0 ... 20 mA

max 60 mA

External loads

current outputs

R

L

< 500 ohms

Max load + cable loop resistance

540 ohms

0 ... .1 V

R

L

> 2000 ohms

0 ... 5 V and 0 ... 10 V

R

L

> 10 000 ohms


Storage temperature range

-55 ... +80 °C (-67 ... +176 ºF)

3-pole screw terminal

max wire size

1.5 mm

2

(AWG16)

Relay Module

Operating temperature range

-40 ... +60 ºC (-40 ... +140 ºF)

Operating pressure range

500 ... 1300 mmHg

Power consumption at 24 V

max 30 mA

Contacts SPDT (change over), for example,
Contact arrangement Form C

Imax

0.5 A 250 VAC

Imax

0.5 A 30 VDC

Safety standard for the relay component

IEC60950 UL1950

Storage temperature range

-55 ... +80 ºC (-67 ... +176 ºF)

3-pole screw terminal / relay

max wire size

2.5 mm

2

(AWG14)

Advertising