Probe mounting, Figure 22, Measurement error at 100 %rh – Vaisala HMT330 User Manual

Page 42

Advertising
background image

User's Guide _______________________________________________________________________

40 ___________________________________________________________________ M210566EN-K

Probe Mounting

In humidity measurement and especially in calibration it is essential that
temperature of the probe and measuring environment is the same. Even a
small difference in temperature between the environment and the probe
causes an error. As the curve below shows, if the temperature is +20 °C
and the relative humidity 100 %RH, a difference of ±1 °C between the
environment and the probe causes an error of ±6 %RH.

The graph below illustrates the measurement error at 100 %RH when the
difference between the ambient and sensor temperature is 1 °C.

0507-023

Figure 22

Measurement Error at 100 %RH

Advertising