User's Guide _______________________________________________________________________

202 __________________________________________________________________ M210566EN-K

use the floating point registers to write values outside this 16-bit signed
integer range. Maximum writable value is 32767 also for configuration
registers that do not accept negative values.

Measurement Data (Read-Only)

Table 41

Measurement Data Registers

Name

Float

Integer

Unit

RH

0001…0002

0257 (×0.01)

%

T

0003…0004

0258 (×0.01)

°C

T

d

0007…0008

0260 (×0.01)

°C

T

d/f

0009…0010

0261 (×0.01)

°C

a

0015…0016

0264 (×0.01)

g/m³

x

0017…0018

0265 (×0.01)

g/kg

T

w

0019…0020

0266 (×0.01)

°C

H

2

O

0021…0022

0267 (×1)

ppm

V

p

w

0023…0024

0268 (×0.1)

hPa

p

ws

0025…0026

0269 (×0.1)

hPa

H

0027…0028

0270 (×0.01)

kJ/kg

ΔT

0031…0032

0272 (×0.01)

°C

H

2

O

0065…0066

0289 (×1)

ppm

W

Available measurements depend on the instrument configuration. Values
may be unavailable also in case of device failure. Read status registers or
exception status outputs to check for failures.

NOTE

Only real-time data can be read from the instrument. Data that is
recorded in the transmitter’s memory (internal memory or data logger
module) cannot be read through Modbus.

Status Registers (Read-Only)

Table 42

Status Registers

Name

Address

Description

Fault status

0513

1 = No errors

Online status

0514

1 = Online data available

Error code (bits 15…0)

0516

Bit field that represents the
error codes listed in Table 33
on page 152. Active errors are
set to 1. For example, if bit 14
is set to 1, error code E14 is
active.

Error code (bits 31…16)

0517


Error information can retrieved also with function code 08, sub-function
02.