Getting measurement message from serial line, Starting continuous outputting, Table 26 – Vaisala HMT330 User Manual

Page 97: Other commands

Advertising
background image

Chapter 4 _________________________________________________________________ Operation

VAISALA ________________________________________________________________________ 95

Table 26

Other Commands

Command

Description

?

Output information about the device

??

Output information about the device in
POLL mode

ALSEL

Set display alarms

CON

Set display contrast

ECHO [ON/OFF]

Turn the serial interface echo ON/OFF

ERRS

List present transmitter errors

FILT

Set the result filtering

FIND

All devices in POLL mode send their
addresses

HELP

List the currently available commands

LIGHT

Set the display backlight mode

LOCK

Lock the menu or disable the keypad

MODBUS

View Modbus diagnostic counters

MODS

View information about connected modules

PRES [hPa]

Set the value for pressure compensations

RHLIMIT [ON/OFF]

Limit highest value of RH output to 100%

VERS

Display the software version information

XHEAT

Sensor heating

XPRES [hPa]

Set the value for pressure compensations,
temporarily

Getting Measurement Message from Serial Line

Starting Continuous Outputting

Use the R command to start the continuous output of measurements.

R<cr>

The default output format depends on the quantities your transmitter
supports. You can change the output message format and interval with
the FORM and INTV commands. See sections FORM on page 101 and
INTV on page 115.

Example:

>r
RH= 60.5 %RH T= 23.7 'C Tdf= 15.6 'C Td= 15.6 'C a= 13.0
g/m3 x= 11.1 g/kg Tw= 18.5 'C H2O= 17889 ppmV pw=
17.81 hPa pws= 29.43 hPa h= 52.3 kJ/kg dT= 8.1 'C

If a value is too long to fit to the allocated space in the output, or if there
is an error in outputting the quantity, the value is displayed with stars ‘*’.

Example:

>r
RH=***.* %RH T= 31.0 'C