Appendix C _________________________________________________________ Modbus Reference

VAISALA _______________________________________________________________________ 205

Code

Sub-Function Name

Notes

13 (0x0D)

Return Bus Exception Error
Count

Number of Modbus exception
responses sent

14 (0x0E)

Return Slave Message Count Number of Modbus messages

processed

15 (0x0F)

Return Slave No Response
Count

Number of Modbus messages
received without sending a
response. This happens for
received broadcast messages
(Modbus RTU only) or when
HMT330 is in the “Listen Only”
mode.


Although Modbus Diagnostics are standardized only for serial line
devices, HMT330 supports the same diagnostic functions also on
Modbus TCP.

NOTE

Resetting, powering up the transmitter, or reselecting the Modbus mode
(by serial command or with the user interface) resets all Modbus
diagnostic counters and cancels any “Listen Only” mode.

Device Identification Objects

HMT330 Modbus conforms to the extended identification level defined
in the Modbus Application Protocol Specification V1.1b. Both stream
access and individual access to the objects is supported.

Table 47

HMT330 Modbus Device Identification

Object Id

Object Name

Description

0x00

VendorName

“Vaisala”

0x01

ProductCode

Product code (e.g. “HMT330”)

0x02

MajorMinorVersion

Software version (e.g. “5.10”)

0x03

VendorUrl

“http://www.vaisala.com/”

0x04

ProductName

Product name of the instrument

0x80

SerialNumber

Serial number (e.g. “D0710040”)

0x81

CalibrationDate

Date of the last calibration
(e.g. “2011-02-07”, empty if not
available)

0x82

CalibrationText

Information text of the last calibration
(empty if not available)