Starting and configuring chemical purge, Using buttons on motherboard, Using display/keypad (optional) – Vaisala HMT330 User Manual

Page 137: Figure 67, Purge buttons on motherboard, Figure 68, Chemical purge settings

Advertising
background image

Chapter 4 _________________________________________________________________ Operation

VAISALA _______________________________________________________________________ 135

Starting and Configuring Chemical Purge

Using Buttons on Motherboard

Start manual chemical purge by pressing simultaneously two

PURGE

buttons on the motherboard inside the transmitter for a few seconds.
Indicator LED flashes until purge is complete (up to 6 minutes).

0508-011

Figure 67

Purge Buttons on Motherboard

Using Display/Keypad (Optional)

1.

Open the Main Menu

by pressing any of the ▼▲◄► arrow

buttons.

2.

Select ►Measuring, press ►button

.

3.

Select ►Chemical purge, press ► button.

1102-016

Figure 68

Chemical Purge Settings

- Start the chemical purge manually by selecting Start purge now

and pressing START.

- Select Purge on power-up by using the arrow buttons. Press

On/Off to turn the start-up purge on/off.

- Turn on/off the automatic purge by selecting Automatic purge,

press ►ON/OFF button.

- Set the automatic purge interval by selecting Interval: ..., press

SET. Set the purge interval and the unit (hour/day) by using the
arrow buttons. The interval must be 1 hour ... 10 days. Press OK.

Advertising