Operating voltage

10 ... 35 VDC, 24 VAC

with optional power supply module

100 ... 240 VAC, 50/60 Hz

Start-up time after power-up

3 s

Power consumption at 20 °C (Uin 24VDC)

RS-232

max 25 mA

Uout 2 × 0 ... 1V / 0 ... 5V / 0 ... 10V

max 25 mA

Iout 2 × 0 ... 20 mA

max 60 mA

display and backlight

+ 20 mA

during sensor purge

+ 110 mA max

Analog outputs (2 standard, 3rd optional)

current output

0 ... 20 mA, 4 ... 20 mA

voltage output

0 ... 1 V, 0 ... 5 V, 0 ... 10 V

Accuracy of analog outputs at 20 °C

± 0.05 % full scale

Temperature dependence of the analog outputs

± 0.005 %/°C full scale

External loads

current outputs

RL < 500 ohm

0 ... 1V output

RL > 2 kohm

0 ... 5V and 0 ... 10V outputs

RL > 10 kohm

Wire size

0.5 ... 2.5 mm

2

(AWG 20 ... 14)

stranded wires recommended

Digital outputs

RS-232

RS-422/485 (optional)

LAN (optional)

WLAN (optional)

Protocols

ASCII commands

Modbus RTU

Modbus TCP

Relay outputs (optional)

0.5 A, 250 VAC, SPDT

Display (optional)

LCD with backlight, graphic
trend display

Menu languages

Chinese, English, Finnish,
French, German, Japanese,
Russian, Spanish, Swedish