Appendix C _________________________________________________________ Modbus Reference

VAISALA _______________________________________________________________________ 203

NOTE

Basic status information of the HMT330 transmitter is also available with
the Read Exception Status function code 07. For details, see section
Exception Status Outputs on page 204.

Configuration Registers

Configuration parameter registers are used to configure the measurement.
Writing out-of-range values is silently ignored.

Table 43

Configuration Parameter Registers

Name

Float

Integer

Valid Range

Pressure setting default
value

0769…0770

1025 (×1)

0…9999 hPa

Pressure setting
temporary value

0771…0772

1026 (×1)

0…9999 hPa

Chemical purge interval 0773…0774

1027 (×1)

10…2880 min

Molecular weight for
ppm

w

calculation

0775…0776

1028 (×0.001)

0…999.999
g/mol


If you modify the pressure setting repeatedly, modify the temporary
value at register 0771...0772 (1026) instead of the permanent setting at
register 0769…0770 (1025). Set the temporary value to zero to return to
the default pressure setting.

Non-metric units are not available on Modbus. If they are needed,
calculate the conversion outside the transmitter.

Configuration flags are used to select some basic options of the
instrument and to manually start the sensor operations.

Table 44

Configuration Flag Registers

Name

Address

Description

Standard filtering on/off

1281

1 = Filtering on

Extended filtering on/off

1282

1 = Extended filtering on

Automatic chemical purge on/off 1283

1 = Automatic purge on.
See configuration register
0773…0774 (1027).

Start-up chemical purge on/off

1284

1 = Start-up purge on

Chemical purge in progress

1285

1 = Purge in progress
(write 1 for manual start)

Sensor heating on/off

1287

1 = Sensor heating on
(optional feature)


Flags 1281 and 1282 are mutually exclusive.

Some of the configuration registers might not have an effect depending
on the instrument model and configuration.