11. When stabilized, type the high end reference value after the

question mark and press ENTER.

>crh
RH : 11.25 Ref1 ? c
RH : 11.24 Ref1 ? c
RH : 11.24 Ref1 ? 11.3
Press any key when ready ...

RH : 75.45 Ref2 ? c
RH : 75.57 Ref2 ? c
RH : 75.55 Ref2 ? c
RH : 75.59 Ref2 ? 75.5
OK
>

12.

OK

indicates that the adjustment has succeeded and the new

calibration coefficients are calculated and stored. Enter the
adjustment information (date and text) to the memory of the
transmitter; see the commands CTEXT and CDATE.

13. Press the ADJ button on the motherboard to close the adjustment

mode.

14. Take the probe out of the reference conditions and replace the

filter.