Temperature (+ operating pressure ranges), Optional temperature probe, Figure 80 – Vaisala HMT330 User Manual

Page 172: Accuracy over temperature range, Figure 80 accuracy over temperature range

Advertising
background image

User's Guide _______________________________________________________________________

170 __________________________________________________________________ M210566EN-K

Temperature (+ Operating Pressure Ranges)

HMT331

-40 ... +60 °C (-40 ... +140 °F)

HMT333 80 ºC

-40 ... +80 °C (-40 ... +176 °F)

HMT333 120 ºC

-40 ... +120 °C (-40 ... +248 °F)

HMT334

-70 ... +180 °C (-94 ... +356 °F),
0 ... 10 MPa (0 ... 100 bar)

HMT335 (vapor tight)

-70 ... +180 °C (-94 ... +356 °F)

HMT337 (vapor tight)

-70 ... +180 °C (-94 ... +356 °F)

HMT338

-70 ... +180 °C (-94 ... +356 °F),
0 ... 4 MPa (0 ... 40 bar)


Accuracy at +20 °C (+68 °F)

± 0.2 °C

Accuracy over temperature range (see graph below):

0507-021

Figure 80

Accuracy over Temperature Range

Temperature sensor

Pt100 RTD Class F0.1 IEC 60751

Optional Temperature Probe

Temperature measurement
range:

-70 ... + 180 ºC (-94 ... +356 ºF)

Typical accuracy:

0.1 ºC (0.18 ºF)

Sensor:

Pt100 PRT DIN IEC 751 class 1/4 B

Cable length:

2 m, 5 m, and 10 m

Pressure tight:

Up to 7 bar

Probe material:

Stainless steel