Appendix A _________________________________ Probe Installation Kits and Installation Examples

VAISALA _______________________________________________________________________ 189

T- Probe Installations (HMT337)

0508-015

Figure 98

Vapor Tight Installation

Vapor Tight Installation is not available from Vaisala.

The following numbers refer to Figure 98:

1

=

Cable gland. For example AGRO 1100.12.91.065
(T= -25 ... +200 ºC)

2

=

In pressurized processes, use a locking ring
(example: 6x 0.7 DIN471)

0508-022

Figure 99

Wall Mounting Installation

Wall Mounting Installation is not available from Vaisala.

The following numbers refer to Figure 99:

1

=

Cable gland. For example AGRO 1100.12.91.065

2

=

Compacted PTFE sleeve

3

=

Silicon glue between the PTFE sleeve and the cable

4

=

Temperature probe

5

=

Recommended support to keep the probe in horizontal
position