Relative Humidity Adjustment after Sensor
Change

Using Display/Keypad

When using the optional display/keypad, follow the instructions in
section Using Display/Keypad on page 158 but select Adj. for new RH
sensor (
instead of 1-point/ 2-point adjustment).

Using Serial Line

After sensor change, carry out the procedure as described in previous
sections. Just replace the CRH command with the FCRH command.

FCRH

Example:


>fcrh
RH

:

1.82 1. ref

?

0

Press any key when ready...

RH

:

74.22

2. ref

? 75

OK
>

The

OK

indicates that the calibration has succeeded.