Calculation formulas, Appendix b, Appendix b, calculation formulas, pr – Vaisala HMT330 User Manual

Page 197: 1log −       = a pw m t t

Advertising
background image

Appendix B _______________________________________________________ Calculation Formulas

VAISALA _______________________________________________________________________ 195

APPENDIX B

CALCULATION FORMULAS

This Appendix contains the formulas used for the calculated output
quantities.

The HMT330 series transmitters measure relative humidity and
temperature. From these values dewpoint, mixing ratio, absolute
humidity and enthalpy in normal pressure are calculated using the
following equations:

Dewpoint:

1

log

=

A

Pw

m

T

T

n

d

(1)


P

w

is the water vapor pressure. The parameters A, m, and T

n

depend on

temperature according to the following table:

t

A

m

Tn

<0 °C

1)

6.1134

9.7911

273.47

0 ... 50 °C

6.1078

7.5000

237.3

50 ... 100 °C

5.9987

7.3313

229.1

100 ... 150 °C

5.8493

7.2756

225.0

150 ... 180 °C

6.2301

7.3033

230.0


1) Used for frostpoint calculation if the dewpoint is negative