Appendix A _________________________________ Probe Installation Kits and Installation Examples

VAISALA _______________________________________________________________________ 187

Pressure Tight Swagelok Installation Kits (for
HMT337)

RH Probe Installation

Swagelok installation kit for the relative humidity probe includes
Swagelok connector with ISO3/8" or NPT1/2" thread. Vaisala order
codes: SWG12ISO38 or SWG12NPT12.

0508-032

Figure 94

Swagelok Installation Kit for RH-probe

The following numbers refer to Figure 94:
1

=

Relative humidity probe

2

=

Duct connector

3

=

ISO3/8" or NPT1/2" thread

4

=

Swagelok connector

5

=

Ferrules

Temperature Probe Installation

Swagelok installation kit for T-probe includes Swagelok connector with
either ISO1/8" or NPT1/8" thread. Vaisala order codes: SWG6ISO18 or
SWG6NPT18.

0508-016

Figure 95

Swagelok Installation Kit for T-Probe

The following numbers refer to Figure 95:
1

=

T-probe

2

=

Duct connector

3

=

Swagelok connector

4

=

Ferrules

135 mm

33 mm

18 mm min

max 30 mm

1 2 3

4

P = max 10 bar
T = max 180 ºC

Advertising